Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
3.0K
The Effect of Low-Dose CT Protocols on Shoulder Model-Based Tracking accuracy Using Biplane Videoradiography
Stacey Chen1, Erin C S Lee2, Kelby B Napier3
1Program in Physical Therapy, Washington University School of Medicine in St. Louis, 4444 Forest Park Avenue, St. Louis, MO, 63108, USA.
Annals of Biomedical Engineering
|November 6, 2024
Summary
Reducing computed tomography (CT) radiation dose for shoulder motion analysis is possible. Low-dose CT scans allow for substantial dose reduction without significantly impacting the accuracy of humerus, scapula, and clavicle kinematic tracking.
Area of Science:
- Orthopedics
- Medical Imaging
- Biomechanics
Background:
- Model-based tracking is crucial for quantifying shoulder kinematics.
- Computed tomography (CT) is commonly used for 3D bone reconstruction, but increases radiation exposure.
- Lower-dose CT protocols may be feasible due to bone-soft tissue contrast.
Purpose of the Study:
- To evaluate the dose-accuracy trade-off between low-dose CT scans and kinematic tracking accuracy.
- To assess the impact of reduced CT dose on the tracking of humerus, scapula, and clavicle.
- To determine the feasibility of using intensity-based registration algorithms with low-dose CT for shoulder motion analysis.
Main Methods:
- Investigated full-dose and four low-dose CT protocols on three fresh-frozen cadavers.
- Modulated X-ray tube current and peak voltage for low-dose protocols.
- Compared model-based tracking results against radiostereometric analysis (reference standard).
- Quantified accuracy using root-mean-square error (RMSE), bias, precision, and worst-case errors for segmental and joint kinematics.
Main Results:
- Achieved average CT dose reductions of 70.6-92.8%.
- Observed slight increases in RMSE for glenohumeral and acromioclavicular joints with decreased dose.
- Found minimal differences (≤0.3°, ≤0.1 mm) in joint kinematic errors between highest and lowest dose protocols.
Conclusions:
- Substantial CT dose reduction is achievable for shoulder motion analysis.
- Low-dose CT protocols do not significantly compromise data fidelity for kinematic tracking.
- Model-based tracking with reduced CT dose offers a viable alternative for shoulder motion analysis.