Substantiating Appropriate Motion Capture Techniques for the Assessment of Nordic Walking Gait and Posture in Older Adults
Published on: May 12, 2016
Establishing the Reliability of the GaitON® Motion Analysis System: A Foundational Study for Gait and Posture
Md Farhan Alam1, Saima Zaki1,2, Saurabh Sharma1
1Centre for Physiotherapy and Rehabilitation Sciences, Jamia Millia Islamia, Maulana Muhammad Ali Jauhar Marg, New Delhi 110025, India.
Background:
Gait and posture analysis plays a crucial role in understanding human movement, with significant applications in rehabilitation, sports science, and clinical settings. The GaitON® system, a 2D motion analysis tool, provides an accessible and cost-effective method for assessing gait and posture. However, its reliability in clinical practice, particularly for intra-rater consistency, remains to be evaluated. This study aims to assess the intra-rater reliability of the GaitON® system in a healthy population, focusing on gait and posture parameters.
Methods:
A total of 20 healthy participants (10 males and 10 females) aged 18 to 50 years were recruited for the study. Each participant underwent gait and posture assessments using the GaitON® system on two separate occasions, spaced one week apart. Video recordings from anterior and posterior views were used to analyze gait, while images from anterior, posterior, and lateral views were captured to assess posture with markers placed on key anatomical landmarks. The reliability of the measurements was analyzed using intraclass correlation coefficients (ICC), a standard error of measurement (SEM), and the smallest detectable difference (SDD) method.
Results:
The GaitON® system demonstrated excellent intra-rater reliability across a wide range of gait and posture parameters. ICC values for gait parameters, including hip, knee, and ankle joint angles, ranged from 0.90 to 0.979, indicating strong consistency in repeated measurements. Similarly, ICC values for posture parameters, such as the head alignment, shoulder position, and ASIS alignment, were above 0.90, reflecting excellent reliability. SEM values were low across all parameters, with the smallest SEM recorded for the hip joint angle (0.37°), and SDD values further confirmed the precision of the system.
Conclusion:
The GaitON® system provides reliable and consistent measurements for both gait and posture analysis in healthy individuals. Its high intra-rater reliability and low measurement error make it a promising tool for clinical and sports applications. Further research is needed to validate its use in clinical populations and compare its performance to more complex 3D motion analysis systems.