Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism?
Arpita Katheria1, Kamlesh Raut1, Ankit Kumar Sahu1
1Department of Cardiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, 226014, Uttar Pradesh, India.
Abstract:
Unexplained change in parameters, despite good lead positions or odd ECG patterns following implant should lead to suspicion of lead swap, amongst other possibilities. An easy way to confirm this is by paying careful attention to device EGMs and recording the ECG by switching off one channel, and pacing single channel (in either AAI or VVI mode).