Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the
Ying Zhang1, Ziyun Shi2, Haixia Liang1
1Department of Obstetrics and Gynecology, Xijing 986 hospital, Fourth Military Medical University, Xi'an, Shaanxi, People's Republic of China.
Objective:
Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).
Methods:
A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.
Results:
The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was -0.84±3.88 mmHg and -0.58±3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.
Conclusion:
The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)