Amplitude spectrum area is dependent on the electrocardiogram magnitude: evaluation of different normalization
Luiz E V Silva1, Hunter A Gaudio2, Nicholas J Widmann2
1Tsui Laboratory, Department of Biomedical and Health Informatics, Children's Hospital of Philadelphia, Philadelphia, PA 19146, United States of America.
Abstract:
Objective.Amplitude Spectrum Area (AMSA) of the electrocardiogram (ECG) waveform during ventricular fibrillation (VF) has shown promise as a predictor of defibrillation success during cardiopulmonary resuscitation (CPR). However, AMSA relies on the magnitude of the ECG waveform, raising concerns about reproducibility across different settings that may introduce magnitude bias. This study aimed to evaluate different AMSA normalization approaches and their impact on removing bias while preserving predictive value.Approach.ECG were recorded in 118 piglets (1-2 months old) during a model of asphyxia-associated VF cardiac arrest and CPR. An initial subset (91/118) was recorded using one device (Device 1), and the remaining piglets were recorded in the second device (Device 2). Raw AMSA and three ECG magnitude metrics were estimated to assess magnitude-related bias between devices. Five AMSA normalization approaches were assessed for their ability to remove detected bias and to classify defibrillation success.Main results.Device 2 showed significantly lower ECG magnitude and raw AMSA compared to Device 1. CPR-based AMSA normalization approaches mitigated device-associated bias. Raw AMSA normalized by the average AMSA in the 1st minute of CPR (AMSA1m-cpr) exhibited the best sensitivity and specificity for classification of successful and unsuccessful defibrillation. While the optimal AMSA1m-cprthresholds for balanced sensitivity and specificity were consistent across both devices, the optimal raw AMSA thresholds varied between the two devices. The area under the receiver operating characteristic curve for AMSA1m-cprdid not significantly differ from raw AMSA for both devices (Device 1: 0.74 vs. 0.88,P= 0.14; Device 2: 0.56 vs. 0.59,P= 0.81).Significance.Unlike raw AMSA, AMSA1m-cprdemonstrated consistent results across different devices while maintaining predictive value for defibrillation success. This consistency has important implications for the widespread use of AMSA and the development of future guidelines on optimal AMSA thresholds for successful defibrillation.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Bode Plots Construction
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....