Performance evaluation of the high-throughput quantitative Alinity m Epstein-Barr virus assay.
D Yitzchak Goldstein1, Mark M Sasaki2, Momka Narlieva1
1Department of Pathology, Montefiore Medical Center, Bronx, New York, USA.
Microbiology Spectrum
|November 15, 2024
Summary
The Alinity m EBV assay accurately quantifies Epstein-Barr virus (EBV) DNA in transplant patients. This molecular test shows high precision and comparable results to existing assays, aiding in PTLD risk management.
Area of Science:
- Virology
- Clinical Diagnostics
- Molecular Biology
Background:
- Epstein-Barr virus (EBV) reactivation poses a significant risk for post-transplant lymphoproliferative disorders (PTLD) in immunocompromised individuals, particularly within the first year post-transplant.
- Quantitative molecular testing for EBV viral load is crucial for monitoring and predicting PTLD risk.
- High-sensitivity and high-specificity nucleic acid amplification tests (NAATs) on automated platforms enhance diagnostic accuracy and turnaround time.
Purpose of the Study:
- To evaluate the analytical and clinical performance of the Alinity m EBV assay for quantitative Epstein-Barr virus detection.
- To assess the assay's precision, limit of detection, and correlation with established EBV testing methods.
Main Methods:
- Analytical performance, including limit of detection and precision, was determined using commercial panels.
- Clinical performance was evaluated by comparing the Alinity m EBV assay results with those from established laboratory-developed tests (LDTs) and a commercial assay (Roche cobas EBV) using remnant patient specimens.
- Statistical analysis included correlation coefficients and mean bias calculations.
Main Results:
- The Alinity m EBV assay demonstrated a limit of detection of 1.3 Log IU/mL and high precision (1.9%–5.2% CV).
- Correlations with reference assays were strong, with coefficients ranging from 0.762 to 0.970.
- Mean bias values indicated comparable performance across different comparison groups.
Conclusions:
- The Alinity m EBV assay exhibits high analytical precision and accuracy across its measurement range.
- Results are comparable to established EBV assays, supporting its utility in clinical settings.
- The Alinity m EBV assay is a valuable tool for managing EBV infection in transplant recipients, aiding in PTLD risk assessment.