Room temperature quantitative liquid concentration device and application to interleukins analysis in a B-cell
Ruriko Kawanabe1, Hidekatsu Tazawa2, Kazuma Mawatari2
1Department of Dermatology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo, 113-8655, Japan.
Abstract:
In biological analysis and medical diagnosis, there is an increasing demand for improving the lower detection limit without deteriorating the quantitativity; however, it is usually challenging. In this study, we utilized a cyclone flow device and established a liquid concentration method. An air cyclone flow induced a liquid cyclone flow in the concentration devices and enhanced the air/liquid interface area, which allowed an effective concentration of liquid from mL to mL at room temperature. The heating temperature and actual temperature inside the liquid were investigated to know the cooling effect of evaporation. The collection efficiency of larger than 98% was confirmed with a standard solution. Finally, the analytical procedure to realize a quantitative concentration was established, and the concentration and quantification of interleukins (IL-8, IL-17, and IL-23) from the supernatant of the B-cell culture medium was demonstrated. The B-cell was stimulated with CD40L, and the supernatant was concentrated 27 times at maximum.