Feasibility study of modularized pin ridge filter implementation in proton FLASH planning for liver stereotactic
Chaoqiong Ma1,2, Xiaofeng Yang1, Jufri Setianegara2
1Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, GA 30322, United States of America.
Abstract:
Objective.We previously developed a FLASH planning framework for streamlined pin-ridge-filter (pin-RF) design, demonstrating its feasibility for single-energy proton FLASH planning. In this study, we refined the pin-RF design for easy assembly using reusable modules, focusing on its application in liver stereotactic ablative body radiotherapy (SABR).Approach.This framework generates an intermediate intensity-modulated proton therapy (IMPT) plan and translates it into step widths and thicknesses of pin-RFs for a single-energy FLASH plan. Parameters like energy spacing, monitor unit limit, and spot quantity were adjusted during IMPT planning, resulting in pin-RFs assembled using predefined modules with widths from 1 to 6 mm, each with a water-equivalent-thickness of 5 mm. This approach was validated on three liver SABR cases. FLASH doses, quantified using the FLASH effectiveness model at 1-5 Gy thresholds, were compared to conventional IMPT (IMPT-CONV) doses to assess clinical benefits.Main results.The highest demand for 6 mm width modules, moderate for 2-4 mm, and minimal for 1- and 5-mm modules were shown across all cases. At lower dose thresholds, the two-beam case reduced indicators including liverV21Gyand skinDmaxby >19.4%, while the three-beam cases showed reductions⩽11.4%, indicating the need for higher fractional beam doses for an enhanced FLASH effect. Positive clinical benefits were seen only in the two-beam case at the 5 Gy threshold. At the 1 Gy threshold, the two-beam FLASH plan outperformed the IMPT-CONV plan, reducing dose indicators for all relevant normal tissues by up to 31.2%. In contrast, the three-beam cases showed negative clinical benefits, with skinDmaxand liverV21Gyincreasing by up to 17.4% due to lower fractional beam doses and closer beam arrangements.Significance.This study evaluated the feasibility of modularizing streamlined pin-RFs in single-energy proton FLASH planning for liver SABR, offering guidance on optimal module composition and strategies to enhance FLASH planning.