Published on: April 26, 2014
Impact of mechanical disturbances to the placement of reference light in Fourier transform interferometers
1Department of Physics, University of Limerick, Limerick V94T9PX, Ireland.
Abstract:
Fourier transform interferometers usually need a dedicated reference light to precisely track optical path difference changes. There are a few different ways to incorporate the reference light into the optical system. This paper conducted experiments to give a test of how different ways of incorporating reference light are affected by mechanical disturbances. It found that placing the reference light in a different path than the test light can be very sensitive to mechanical disturbance errors, while propagating the reference light together with the test light can be very robust against mechanical vibrations of optical components.
Related Concept Videos
IR Spectrometers
Interference: Path Lengths
Two special sources may be considered when they are in phase. This can be easily achieved by feeding the two sources from the same source. An example would be synchronizing the two speakers by feeding them with the same source, such as the sound waves produced by a tuning fork. This setup ensures that the two sources have the same frequency and are...