Related Experiment Video
Updated: May 17, 2026

09:31
Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
11.7K
Validation of an Augmented Reality Based Functional Method to Determine and Render the Hip Rotation Centre During
Quentin Neuville1,2, Taylor Frantz3,4, Frederick Van Gestel1,2
1Department of Neurosurgery, Vrije Universiteit Brussel, Universitair Ziekenhuis Brussel, Brussels, Belgium.
Summary
This study introduces an augmented reality system for precisely locating the hip's center of rotation during total hip replacement surgery. The method ensures accurate and repeatable measurements for improved surgical outcomes.
Area of Science:
- Orthopedic Surgery
- Medical Imaging
- Augmented Reality Technology
Background:
- Accurate determination of the functional center of rotation (FCOR) is crucial in total hip arthroplasty.
- Existing methods for FCOR determination can be complex and time-consuming.
Purpose of the Study:
- To present a novel method for determining and visualizing the hip's FCOR using an augmented reality head-mounted display (AR-HMD).
- To evaluate the accuracy and reproducibility of the proposed AR-HMD system in an experimental setting.
Main Methods:
- Developed custom software for a HoloLens AR-HMD to track markers on cadaver femurs and 3D-printed acetabuli.
- Collected 80 measurements by rotating 20 cadaver femurs twice in a matching cup.
- Calculated FCOR based on point cloud data derived from marker displacement relative to acetabular trackers.
Main Results:
- The AR-HMD system achieved an absolute error of 2.9 ± 1.4 mm for the acetabulum and 2.9 ± 1.2 mm for the femur.
- 95th percentile errors were below 5.6 mm for the acetabulum and 4.7 mm for the femur, indicating high precision.
- Demonstrated the system's capability in an experimental setting with cadaveric specimens.
Conclusions:
- The proposed AR-HMD system provides an accurate and reproducible method for determining both femoral and acetabular FCOR.
- This technology has the potential to enhance precision in total hip arthroplasty procedures.
- Further clinical validation is warranted to translate these experimental findings into surgical practice.