Optimizing Arterial Tissue Thickness Measurement Protocols: Digital Vernier Caliper Versus Digital Thickness Gauge
Alexandru Petru Ion1,2, Alexandra Asztalos3, Claudiu Constantin Ciucanu1,4,5
1Doctoral School of Medicine and Pharmacy, George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, 540139 Targu Mures, Romania.
Background:
The aim of this study is to analyze the reproducibility of sample thickness measurements taken by a non-experienced user by comparing a standard digital vernier caliper, with four different protocols, to a specialized thickness gauge.
Methods:
The current study is a methodological study where we examined the thickness of the porcine arterial wall in the thoracic aorta of six pigs. Two adjacent samples of 10 × 10 mm from each aorta were excised longitudinally from the anterior wall, resulting in twelve specimens. Five protocols were employed to measure the thickness of each sample. In four of these protocols, digital vernier calipers (Multicomp PRO MP012475) were utilized, while the fifth protocol utilized a specialized digital thickness gauge (Mitutoyo 547-500S, Mitutoyo Corp., Kawasaki, Japan).
Results:
We observed a higher average thickness of the samples during the initial measurement compared to the second measurement (1.11 ± 0.16 vs. 0.94 ± 0.17, p = 0.0319) with the first protocol and smaller values than those determined at the last measurement (0.93 ± 0.15 vs. 1.10 ± 0.15, p = 0.0135) for the third protocol. Further, with the digital vernier calipers, we recorded lower values for all four protocols than for the digital thickness gauge determinations. In addition, we computed the ratio of the thicknesses measured during the first, second, and third measurements to analyze how consistent the values were across the three consecutive measurements, with no difference regarding the third, fourth, and control protocols.
Conclusions:
The digital thickness gauge offers dependable measurements, regardless of the user's expertise in assessing tissue thickness, and demonstrates a substantially higher reproducibility when compared to the digital vernier. We also found that taking an average of the thickness measurements from four specific points on each half of the sides or on each diagonal of each corner yielded consistently reliable results over time when using a standard digital vernier caliper instead of a specialized one.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure