Next generation tomosynthesis image acquisition optimization for dedicated PET-DBT attenuation corrections.
Trevor L Vent1, Bruno Barufaldi1, Raymond J Acciavatti1
1Department of Radiology, University of Pennsylvania, Philadelphia, United States.
Next-generation tomosynthesis (NGT) with integrated positron emission tomography (PET-DBT) shows improved dense tissue classification and breast segmentation. Optimized acquisition geometries enhance image quality and reduce volume overestimation compared to conventional digital breast tomosynthesis.
Area of Science:
- Medical Imaging
- Radiology
- Biomedical Engineering
Background:
- Digital breast tomosynthesis (DBT) is a key imaging modality for breast cancer detection.
- Integrating positron emission tomography (PET) with DBT (PET-DBT) offers potential for improved diagnostic accuracy.
- Optimizing acquisition geometries is crucial for enhancing PET-DBT performance.
Purpose of the Study:
- To investigate custom acquisition geometries for a next-generation tomosynthesis (NGT) prototype.
- To improve dense/adipose tissue classification within the breast.
- To enhance breast outline segmentation for better image analysis.
Main Methods:
- Simulated various NGT acquisition geometries using the OpenVCT virtual clinical trial framework.
- Employed anthropomorphic breast phantoms for realistic simulations.
- Compared custom NGT geometries (40-200 mm PA motion) against conventional DBT.
- Utilized receiver operating characteristic (ROC) curves and area under the curve (AUC) to quantify classification performance.
- Calculated reconstructed breast volume and sensitivity index (d').
Main Results:
- Increasing posteroanterior (PA) x-ray source motion in NGT reduced breast volume overestimation compared to conventional DBT.
- Both AUC and sensitivity index (d') increased with greater PA source motion.
- Custom NGT geometries demonstrated superior adipose-glandular classification and segmentation.
Conclusions:
- Next-generation tomosynthesis (NGT) acquisition geometries can significantly improve breast tissue classification and segmentation.
- Optimized NGT parameters enhance image quality for PET-DBT applications.
- These findings suggest NGT offers improved attenuation corrections for PET-DBT over conventional DBT.