Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Mechanical studies for the rotational stability of a cemented stem in cases with stem anteversion adjustment in the
Ariha Goshi1, Yu Takeda2, Takuya Nakai3
1Department of Orthopedic Surgery, Goshi Hospital, 10-8-20, Nagasunishidoori, Amagasaki City, Hyogo, 660-0807, Japan.
Introduction:
The cemented stem can accommodate various shapes of femurs and is a stable stem with long-term results. Additionally, the cemented stem is a highly versatile stem that can be used to adjust stem anteversion in patients with variable femoral anteversion. On the other hand, rotating a stem with increasing cement viscosity may result in reduced postoperative rotational stability.
Materials And Methods:
Rotational stability was compared using two different cementing techniques via in vitro models. Six stems in Group S (Static) were inserted and fixed until the cement was completely hardened. Six stems in Group R (Rotation) were inserted and then rotated 15° before the cement had fully cured. After the cement had completely set, a static rotation testing was performed in which the stem was rotated at a rate of 1°/sec to 20° while a vertical load of 600 N was applied. For the dynamic testing, a vertical load of 600 N and 10,000 cycles of repetitive rotation were applied at 10° around the stem axis at 0.1 Hz. After 10,000 cycles, maximum torque was compared between Groups S and R.
Results:
In both groups, the torque value increased proportionally to the rotation angle. There were no significant differences in resistance to static rotation between Groups S and R (P = 0.599). In the dynamic testing, there was no significant difference between Groups S and R in both minimum torque and maximum torque in any of the cycles. There were no differences in the rotational torque between Groups S and R at 10,000 cycles (P = 0.693).
Conclusions:
The rotational stability of the cemented stems did not decrease in vitro when stem version control was performed up to 15° before the cement had completely hardened.
More Related Videos
08:15Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Residual Stresses in Circular Shafts
Transformation of Plane Stress
Angle of Twist: Problem Solving
Stability of structures
Soundness of Cement
Deformation in a Circular Shaft