Comparison of Procedural Outcomes of Lumenless Fixed-Helix Versus Stylet-Driven Extendable-Helix Lead Systems in Left
Jenish P Shroff1,2, Anugrah Nair1,2, Deep Chandh Raja1
1The School of Medicine and Psychology, Australian National University, Australian Capital Territory (J.P.S., A.N., D.C.R., W.P.A., R.K.P.).
Background:
Left bundle branch pacing (LBBP) has emerged as a safe and effective alternative to right ventricular pacing. Traditionally, LBBP is performed with lumenless lead (LLL); however, the use of stylet-driven lead (SDL) is on rise. We aimed to assess acute success and procedural outcomes of SDL versus LLL for LBBP.
Methods:
One hundred consecutive patients with bradyarrhythmia, indication of cardiac resynchronization therapy, or ablate and pace strategy were randomized in a 1:1 fashion to the SDL and LLL arms. Tendril STS lead with a CPS Locator 3D catheter and SelectSecure 3830 lead with a C315HIS catheter were used in the SDL and LLL arms, respectively. LBBP was confirmed by standard criteria with measurements done on Labsystem Pro.
Results:
Patients in the LLL arm were significantly younger (71.9±11 versus 76.4±8.9 years; P=0.02); all other baseline characteristics were not significantly different. Acute success in LBBP was similar with SDL versus LLL (90% versus 92%; P=0.7). In patients with successful LBBP, screw attempts were not significantly different between the groups (2.3±1.7 in SDL versus 1.9±1.3 in LLL; P=0.2). Implant duration (11±9.6 versus 9.9±7.1 minutes; P=0.4), mean fluoroscopy dose (65.3±82.7 versus 53.5±50.5 mGy; P=0.5), and fluoroscopy time (7.8±4.8 versus 7.4±4 minutes; P=0.7) were also not different in the SDL versus the LLL arm, respectively. Incidence of lead failure (P=0.6), microdislodgement (P=1), and macrodislodgement (P=0.6) were not significantly different. Pacing threshold was comparable at implant and on follow-up at 1, 3, and 6 months.
Conclusions:
LBBP was feasible with both lead systems with similar success rate and low capture threshold. No significant difference was observed in procedure duration or fluoroscopy use. No major complications were recorded with either lead.
Registration:
URL: https://www.anzctr.org.au; Unique identifier: ACTRN12624000304538.
Related Concept Videos
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played by the...
Frames: Problem Solving I
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Circular Shaft - Stresses in Linear Range
Euler's Formula to Columns: Problem Solving
The system comprises two vertical rigid bars, AB and BC, of...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...