Multicenter Clinical Performance Evaluation of the NeuMoDx CT/NG Assay 2.0.
Barbara Van Der Pol1, Ann Avery2, Stephanie N Taylor3
1From the University of Alabama at Birmingham, Heersink School of Medicine, Birmingham, AL.
The NeuMoDx CT/NG Assay 2.0 shows high accuracy for detecting Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) infections using various specimen types. This molecular assay offers flexible throughput options for laboratories, improving access to STI testing.
Area of Science:
- Clinical diagnostics
- Molecular biology
- Public health
Background:
- Rising rates of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) infections necessitate improved diagnostic methods.
- The NeuMoDx CT/NG Assay 2.0 was developed to address the need for accessible sexually transmitted infection (STI) testing.
Purpose of the Study:
- To evaluate the clinical performance of the NeuMoDx CT/NG Assay 2.0.
- To assess the assay's accuracy across different specimen types and platforms.
Main Methods:
- Clinical sensitivity and specificity were determined using endocervical swabs, vaginal swabs (self- and clinician-collected), and first-catch urine specimens.
- Assay results were compared against established US Food and Drug Administration-cleared diagnostic tests.
- Performance was evaluated on both the NeuMoDx-96 (moderate-throughput) and NeuMoDx-288 (high-throughput) molecular systems.
Main Results:
- The NeuMoDx CT/NG Assay 2.0 exhibited high sensitivity and specificity for both CT and NG across most specimen types (≥95% sensitivity, >99% specificity).
- Endocervical sample sensitivity was 93.2% for CT and 93.3% for NG.
- Invalid result rates were low (<1%), with no performance differences observed between the NeuMoDx-96 and NeuMoDx-288 platforms.
Conclusions:
- The NeuMoDx CT/NG Assay 2.0 provides performance comparable to existing FDA-cleared assays.
- The assay's availability on moderate- and high-throughput platforms offers flexibility for laboratories of varying volumes.
- This assay can enhance STI testing capacity, potentially reducing reliance on external reference laboratories.