Abstract:
We describe a method for determining glucose in serum with glucose dehydrogenase immobilized on the inner walls of glass tubes. The reactor was incorporated into a channel of a continuous-flow analyzer (the Technicon SMAC) and used daily for four weeks in routine analysis for serum glucose. During this period we tested the linearity, precision, accuracy, and durability of the reactor. Results correlated well (r = 0.9949) with those obtained by routine methods with free (nonimmobilized) glucose dehydrogenase. The method is shown to be practicable for use in the routine laboratory.