An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Assessment of Efficacy of Implant Screwdrivers' During Multiple Tightening/ Untightening Cycles. An in vitro Study
R Batista1, A Moreira1,2, N Alves3
1Faculdade de Medicina Dentária, Universidade do Porto - R. Dr. Manuel Pereira da Silva, 4200- 393, Porto, Portugal.
Introduction:
It is unclear the number of times a screwdriver may be reused without loosening its efficacy or inducing damage to implant screws while keeping appropriate torque values.
Objective:
Evaluate the effect of the screwdriver and internal threads on the tightening efficacy by assessing the reverse torque value (RTV).
Methods:
The efficacy was tested with four groups of screws: two titanium groups and two gold-platted (n=5). Each screw was tightened 10 times at 20 Ncm using the screwdriver attached to ISD900 (NSK®). In the first protocol one implant was used for the five screws. In the second protocol for each screw a new implant was used. At each untightening, the RTV was measured and recorded.
Results:
RTVs for Gold were statistically lower compared to the Titanium. The RTVs differed among the tested screws. In the first protocol, the first serie tightening/untightening differed statistically from the last three series.
Conclusion:
Titanium screws from Zimmer Biomet achieve better RTV than gold-platted. By increasing the number cycles it is expected to have a decrease in the RTV. The extensive use of screwdrivers and the internal threads of the dental implants might be factors influencing the loss of RTV.