Quantitative sizing of microplastics up to 20 µm using ICP-TOFMS
Fazzolari Sandro1, Hattendorf Bodo1, Günther Detlef1
1ETH Zurich, Department of Chemistry and Applied Biosciences Vladimir-Prelog-Weg 1 8093 Zurich Switzerland guenther@inorg.chem.ethz.ch.
Journal of Analytical Atomic Spectrometry
|December 12, 2024
Summary
This study determined the maximum microplastic size measurable by ICP-TOFMS. Specialized sample introduction systems extended the detection limit to 20 µm for microplastic particles.
Area of Science:
- Environmental Science
- Analytical Chemistry
- Materials Science
Background:
- Accurate quantification of microplastics (MPs) is crucial for environmental monitoring.
- Inductively coupled plasma-time-of-flight-mass spectrometry (ICP-TOFMS) is a powerful tool for elemental analysis, but its application to MP sizing has limitations.
- The upper size limit for MP detection using ICP-TOFMS needs to be better understood and potentially extended.
Purpose of the Study:
- To evaluate the upper size limit of microplastics measurable by ICP-TOFMS.
- To investigate the effectiveness of different sample introduction systems for extending this limit.
- To validate particle sizing using ICP-TOFMS measurements.
Main Methods:
- Tested four sample introduction systems with certified microplastic standards (polystyrene, PMMA, PVC) in suspension (3-20 µm).
- Compared pneumatic nebulization with a falling-tube device and a vertical downwards-pointing ICP-TOFMS.
- Quantified particle size using dissolved citric acid and analyzed chlorine signals for PVC particles.
Main Results:
- Pneumatic nebulization detected microplastics up to 10 µm.
- Falling-tube device and vertical ICP-TOFMS extended the upper size limit to 20 µm.
- Critical size values determined were 2.3 µm (PS), 2.4 µm (PMMA), 3.0 µm (PVC), and 3.9 µm (PVC via chlorine signal).
Conclusions:
- Sample introduction systems significantly impact the upper size limit for microplastic analysis by ICP-TOFMS.
- Falling-tube and vertical ICP-TOFMS methods enable the detection of larger microplastics up to 20 µm.
- ICP-TOFMS is a viable technique for sizing microplastics, with critical size values established for tested polymers.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
632
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
632
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
386
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
386