Achieving reduced radiation exposure with maintained fluoroscopy effectiveness using ultralow-dose settings in
Takashi Kaneshiro1, Sadahiro Murota1, Takeshi Nehashi1
1Department of Cardiovascular Medicine Fukushima Medical University Fukushima Japan.
Journal of Arrhythmia
|December 13, 2024
Summary
Reducing radiation exposure during cryoballoon pulmonary vein isolation (CB-PVI) is crucial. This study found that ultralow dose settings and avoiding cine acquisition significantly decrease radiation doses without affecting procedure time.
Area of Science:
- Medical Imaging
- Cardiology
- Radiation Oncology
Background:
- Radiation exposure during cryoballoon pulmonary vein isolation (CB-PVI) requires optimization.
- Investigating different X-ray system settings is essential for reducing patient radiation doses.
Purpose of the Study:
- To compare radiation doses among three X-ray system settings during CB-PVI.
- To evaluate the impact of anti-scatter grids and dose settings on radiation exposure.
Main Methods:
- 148 patients undergoing first-time CB-PVI were prospectively enrolled into three groups: low dose with grid (LD+G), low dose without grid (LD-G), and ultralow dose with grid (ULD).
- Radiation doses (cumulative air Kerma) were compared across groups.
- A subset of 27 patients underwent CB-PVI using fluoroscopy only (no cine acquisition) for further dose comparison.
Main Results:
- Median cumulative air Kerma was significantly lower in the LD-G (71.8 mGy) and ULD (73.0 mGy) groups compared to LD+G (145.0 mGy).
- For patients without cine acquisition, median cumulative air Kerma was further reduced in LD-G (31.4 mGy) and ULD (22.7 mGy) groups versus LD+G (64.6 mGy).
- Procedure and fluoroscopy times showed no significant differences among groups.
Conclusions:
- Ultralow dose (ULD) X-ray settings combined with avoiding cine acquisition can effectively reduce radiation exposure during CB-PVI.
- These strategies maintain necessary fluoroscopy time, ensuring procedural safety and efficacy.
Keywords:
atrial fibrillationcryoballoon pulmonary vein isolationradiation exposureultralow‐dose setting