Characterizing the Effect of Simultaneous Enhancements of Reducing Gas Species on Figaro Taguchi Gas Sensor
Adil Shah1, Olivier Laurent1, Grégoire Broquet1
1Laboratoire des Sciences du Climat et de l'Environnement (CEA-CNRS-UVSQ), Institut Pierre-Simon Laplace, Université Paris-Saclay, Site de l'Orme des Merisiers, Gif-sur-Yvette 91191, France.
Abstract:
The resistance of the Figaro Taguchi Gas Sensor (TGS) decreases when exposed to reducing gas enhancements. TGS gas response can be characterized by comparing measured resistance to a reference resistance, representative of sampling in identical environmental conditions but with no reducing gas enhancement. Thus, this resistance ratio (RR) allows for characterization of reducing gas response, independent of other environmental effects. This work presents controlled laboratory experiments, measurements, and modeling for an analysis on the effect of reducing gas cross-sensitivities on RR. The methane mole fraction ([CH4]) was raised to approximately 9 ppm from a 0.492 ppm reference level, and carbon monoxide mole fraction ([CO]) was raised to approximately 4 ppm from a 0 ppm reference level, through multiple simultaneous steps. The independent effect of each gas on RR was directly multiplied, resulting in an inferior RR compared with measurements, implying an interdependence effect. For example, for one TGS unit, when deriving [CH4] from RR, a 6 ppm [CH4] measurement would be underestimated by 6% at 1 ppm [CO], but only by 1.6% at 0.1 ppm [CO]. A key implication of residual interdependence effects is that any gas characterization must be conducted with the same reference levels of each other reducing gas expected during field deployment, even if measuring a single gas. A first-order interdependence correction is proposed to account for such interdependence effects. Yet, each TGS behaves differently, and interdependence testing takes time. Therefore, the TGS best serves to detect single reducing gases, assuming all other reducing gases to remain constant at their reference levels.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Atomic Emission Spectroscopy: Interference