Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged
Sajjad Raghavi1, Hamid-Reza Sadoughi2, Mohammad Ehsan Ravari1,3
1Department of Medical Physics, Faculty of Medicine, Semnan University of Medical Sciences, Semnan, Iran.
Background:
Different dose calculation methods vary in accuracy and speed. While most methods sacrifice precision for efficiency Monte Carlo (MC) simulation offers high accuracy but slower calculation. ISOgray treatment planning system (TPS) uses Clarkson, collapsed cone convolution (CCC), and fast Fourier transform (FFT) algorithms for dose distribution. This study's primary goal is to evaluate the dose calculation accuracy for ISOgray TPS algorithms in the presence of a wedge.
Methods:
This study evaluates the dose calculation algorithms using the ISOgray TPS in the context of radiation therapy. The authors compare ISOgray TPS algorithms on an Elekta Compact LINAC through MC simulations. The study compares MC simulations for open and wedge fields with ISOgray algorithms by using gamma index analysis for validation.
Results:
The percentage depth dose results for all open and wedge fields showed a more than 98% pass rate for points. However, there were differences in the dose profile gamma index results. Open fields passed the gamma index analysis in the in-plane direction, but not all points passed in the cross-plane direction. Wedge fields passed in the cross-plane direction, but not all in the in-plane direction, except for the Clarkson algorithms.
Conclusion:
In all investigated algorithms, error increases in the penumbra areas, outside the field, and at cross-plane of open fields and in-plane direction of wedged fields. By increasing the wedge angle, the discrepancy between the TPS algorithms and MC simulations becomes more pronounced. This discrepancy is attributed to the increased presence of scattered photons and the variation in the delivered dose within the wedge field, consequently impacts the beam quality. While the CCC and FFT algorithms had better accuracy, the Clarkson algorithm, particularly at larger effective wedge angles, exhibited greater effectiveness than the two mentioned algorithms.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Design Example: Aggregate Gradation
The grading, or particle-size distribution, of sand is determined using sieve analysis, with standard sizes ranging from 150 μm to 10 mm (ASTM No. 100 sieve to 3⁄8 in. sieve). Sand is sampled...
Pipe Flowrate Measurement: Problem Solving
Area Problem