Comparing Refrigeration to Immediate Room Temperature Testing for Uric Acid Monitoring in Rasburicase-Treated
Leo Lin1,2, Michael Filtz3, Jeffrey Wilson2,4
1Department of Pathology, University of Utah, Salt Lake City, UT, United States.
The Journal of Applied Laboratory Medicine
|December 18, 2024
Summary
Specimens for uric acid monitoring can be tested at room temperature (RT) if analyzed within 45 minutes. Delaying testing of RT samples can lead to significant decreases in uric acid levels, impacting accuracy.
Area of Science:
- Clinical Chemistry
- Preanalytical Variables
- Pharmacokinetics
Background:
- Rasburicase maintains activity at room temperature (RT), necessitating cooling for uric acid monitoring specimens.
- Current preanalytical requirements for uric acid testing may hinder compliance.
Purpose of the Study:
- To evaluate if easing preanalytical requirements for uric acid specimen transport and storage improves compliance without compromising accuracy.
- To determine the impact of room temperature versus iced transport on uric acid stability.
Main Methods:
- Fifty pairs of patient specimens were collected for uric acid-level monitoring.
- Specimens were transported and stored either on ice or at room temperature.
- Uric acid levels were tested at three time points: immediately (approx. 45 min), 90 min, and 135 min postcollection.
Main Results:
- Uric acid concentrations showed no significant clinical difference between room temperature and iced specimens when tested within 45 minutes.
- A negative bias was observed in room temperature specimens tested at 90 min (-9.1%) and 135 min (-17.5%).
- Specimens tested 90 minutes postcollection at room temperature showed an additional 24% decrease in uric acid if tested within two rasburicase half-lives; higher rasburicase doses also influenced results.
Conclusions:
- Rapid laboratory testing of uric acid specimens (within 45 minutes) may allow for validation of alternative preanalytical methods, moving away from ice transport.
- Laboratories capable of prompt testing can potentially relax stringent cooling protocols for uric acid monitoring.