Automated and quantitative assessment of aortic root based on cardiac computed tomography angiography using a new
He Zou1,2,3, Yiqiu Jiang1,2, Haorui Huang4
1Zhejiang-Ireland Joint Laboratory for Precision Diagnosis and Treatment of Valvular Heart Diseases, The Second Affiliated Hospital of Wenzhou Medical University, Wenzhou, China.
Background:
Accurate assessment of aortic root is crucial for the preprocedural planning of transcatheter aortic valve replacement (TAVR). A variety software is emerging for the semiautomated or automated measurements during TAVR planning. This study evaluated a new deep-learning (DL) tool based on cardiac computed tomography angiography (CCTA) for fully automatic assessment of aortic root.
Methods:
The study included 126 patients with CCTA, 63 of whom underwent TAVR. In the overall population, the DL method was compared to manual measurements of the annulus dimensions. Within the TAVR group, the DL method was also compared to 3mensio software-derived aortic root measure, including the annulus, left ventricular outflow tract (LVOT), sinotubular junction (STJ), ascending aorta (AAo), and the heights of both the coronary ostia.
Results:
Data were successfully analyzed using the DL method in 122 (96.8%) of patients. The correlation of annular diameters between the DL and manual methods was good to excellent for the overall cohort (n=118; r=0.83), the TAVR group (n=59, r=0.86), and its subgroups [bicuspid aortic valve (BAV): n=12, r=0.74; tricuspid aortic valve (TAV): n=47, r=0.93]. In the comparison of the DL method with 3mensio, the highest correlation was found for AAo (r=0.99). Among the four diameter indices [minimum, maximum, perimeter-derived diameter (pDD), and area-derived diameter (aDD)], excellent correlation was observed for aDD (LVOT: r=0.92; annulus: r=0.89).
Conclusions:
The DL method offers an effective and efficient tool for the quantification of aortic roots for TAVR planning.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT