Biomechanical Evaluation of Clival Screw Fixation for Occipitocervical Instablity: A Finite Element Analysis
Weipeng Lin1, Jianying Zheng2, Meichao Zhang3
1Department of Orthopaedics, Yunfu People's Hospital, Yunfu, China.
Objective:
The clivus is trapezoidal in shape with uneven bone structure, the optimal number and position of screws for clival fixation are not clear. Therefore, this study aims to explore the optimization clival screw fixation method for occipitocervical instability using finite element analysis.
Methods:
Seven finite element models were developed to evaluate biomechanical properties of clival screw fixation for treating occipitocervical stability, including (i) one clival screw fixation A1 and A2 models; (ii) two clival screws fixation B1 and B2 models; (iii) three clival screws fixation C1 and C2 models; (iv) four clival screws fixation D1 model. Loads of 1.5 Nm were applied to the model fRoM different directions to induce flexion, extension, lateral bending, and axial rotation movements.
Results:
The regular triangle C1 type three clival screws fixation exhibited great stability, with RoM of 4.20° in flexion, 5.80° in extension, 0.85° in lateral bending, and 1.60° in axial rotation. The peak stress on the internal fixation devices were relatively low, with maximum screw stress of 194 MPa in flexion, 276 MPa in extension, 180 MPa in lateral bending, and 213 MPa in axial rotation; the maximum plate stress were 126, 554, 426, and 378 MPa, respectively. The areas with higher stress were mainly concentrated at the robust neck section of the plate.
Conclusion:
The triangular configuration of three clival screws fixation represented the optimized anterior occipitocervical fixation method through the clivus, offering superior biomechanical stability, lower stress on the devices and dispersed stress distribution in the occipitocervical region.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Eccentric Axial Loading in a Plane of Symmetry
Stability of structures