1Precision Medicine - Biomarker & Bioanalytical Platforms (BBP), GSK, Collegeville, PA, USA.
Bioanalysis
|December 20, 2024
Summary
The MSD S-plex platform demonstrated superior sensitivity for cytokine quantification compared to Olink Target 48 and Quanterix SP-X. Fit-for-purpose assessment is crucial for selecting the best cytokine assay for drug development.
Area of Science:
- Biotechnology
- Assay Development
- Immunology
Background:
- Quantifying cytokines is critical in biological research and drug development.
- Numerous assay platforms exist, making optimal selection challenging due to performance variability.
Purpose of the Study:
- To compare the performance of three sensitive, multiplex cytokine assays: MSD S-plex, Olink Target 48, and Quanterix SP-X.
- To benchmark these platforms against the widely used MSD V-plex assay.
Main Methods:
- Serum and plasma samples were analyzed across all platforms.
- Analyte quantification, sample quantifiability, and correlation were assessed.
- MSD S-plex specificity was validated using parallelism and antibody knockdown experiments.
Main Results:
- MSD S-plex exhibited the highest sensitivity, followed by Olink Target 48, Quanterix SP-X, and MSD V-plex.
- While concentrations varied between platforms, strong correlations were observed.
- MSD S-plex results were validated through specificity experiments.
Conclusions:
- MSD S-plex is recommended as a priority platform for ultra-sensitive cytokine assays.
- Olink Target 48 offers a valuable combination of sensitivity and multiplexing for broad cytokine analysis.
- All evaluated platforms (MSD V-plex, MSD S-plex, Olink) are useful in drug development, but performance must be validated per analyte.