Direct Implementation of MSn Using Frequency Scanning Collision Induced Dissociation in a Digital Ion Trap Mass
Jie Ren1, Pingping Chen1, Yunjing Zhang1
1School of Electronic and Information Engineering, Soochow University, Suzhou 215006, China.
Abstract:
Tandem mass spectrometry (MSn) is one of the most effective methods to obtain the structures of organic molecules, enabling the observation of multigenerational ion fragments. Collision-induced dissociation (CID) is currently the most mature technique for mass spectrometry analysis. Ion trap mass spectrometry (ITMS) is favored for on-site detection field, due to its ability of MSn analysis with a single trap and its small size. However, conventional MSn analysis in ITMS requires repeated isolation and excitation processes multiple times, causing high complexity of the entire scanning process. Additionally, the fragment ion detection in ITMS is limited by low-mass cutoff (LMCO) and the weak fragmentation yield. In this study, a method named reverse scanning-collision induced dissociation (RS-CID) is proposed, which involves increasing RF and AC frequencies while maintaining RF voltage constant during the CID process. Twelve representative illegal drugs were analyzed adopting this method, enhancing the intensities of low-mass fragment ions compared to conventional dissociation method. Moreover, experimental results with ketamine and methamphetamine show that RS-CID effectively reduces the LMCO effect and slightly improves CID efficiency. It also enables direct acquisition of their multigeneration fragment ions spectra in a single sequence of ion injection, cooling, isolation, RS-CID, cooling, mass scanning and empty. The experiments to distinguish between the isomers ab-4en-pinaca and adb-3en-butinaca as well as the isomeric compounds 5f-cumyl-pegaclone and cumyl-pipetinaca were also successful by this method. In summary, RS-CID enables MSn analysis in a single sequence and improves the low-mass fragment ions intensity. It can simplify workflows, achieve faster analysis and provide more valuable mass spectral information.
More Related Videos
Related Concept Videos
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Mass Analyzers: Common Types
Mass Spectrometers
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...