Visual, Refractive, Functional, and Patient Satisfaction Outcomes After Implantation of a New Extended Depth-of-Focus
Javier Ferrando Gil1, Adela Churruca Irazola1, Iraia Reparaz1
1Department of Ophthalmology, Donostia University Hospital, Donostia-San Sebastián, Spain.
Clinical Ophthalmology (Auckland, N.Z.)
|December 23, 2024
Summary
The Bi-Flex POB-MA 877PEY extended depth-of-focus intraocular lens (EDoF IOL) offers good distance vision and adequate intermediate vision. Patient satisfaction was high, with minimal visual disturbances like glare or halos reported.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optics
Background:
- Cataract surgery often involves intraocular lens (IOL) implantation to restore vision.
- Extended depth-of-focus (EDoF) IOLs aim to provide a wider range of clear vision compared to traditional monofocal IOLs.
- Evaluating the performance of new EDoF IOL designs is crucial for improving patient outcomes.
Purpose of the Study:
- To evaluate the visual, refractive, functional, and patient satisfaction outcomes of the Bi-Flex POB-MA 877PEY EDoF IOL.
- To assess the efficacy and safety of this specific EDoF IOL model in cataract surgery patients.
Main Methods:
- Prospective longitudinal descriptive study involving 10 patients (20 eyes) implanted with the Bi-Flex POB-MA 877PEY EDoF IOL.
- Monocular visual acuity (distance, intermediate, near) measured at 1 and 3 months post-operatively.
- Binocular contrast sensitivity, defocus curve, and patient-reported satisfaction (VFQ) assessed at 3 months.
Main Results:
- Achieved good uncorrected visual acuity at distance (CDVA: 0.05±0.06 LogMAR at 3 months) and adequate intermediate vision (CIVA: 0.29±0.08 LogMAR at 1 month).
- Refractive outcomes were stable, with 100% of eyes within ±0.5 D at 1 and 3 months post-op.
- High patient satisfaction reported, with 90% experiencing no/minor glare/flare and 100% no/minor halos; 80% rated overall vision 8-9/10.
Conclusions:
- The Bi-Flex POB-MA 877PEY EDoF IOL demonstrates favorable visual and refractive outcomes for distance and intermediate vision.
- The IOL provides a good safety profile with low incidence of visual disturbances.
- This EDoF IOL represents a viable option for cataract surgery patients seeking improved functional vision.
Related Concept Videos
Focusing of Light in the Eye
2.5K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
2.5K
Glaucoma: Overview
505
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
505
Angle Closure Glaucoma: Treatment
435
Angle-closure glaucoma, or closed-angle glaucoma, is an eye condition where the iris bulges out and blocks the iridocorneal angle, resulting in a buildup of aqueous humor and increased intraocular pressure. Immediate medical attention is necessary due to the sudden onset of symptoms. The treatment for angle-closure glaucoma includes short-term and long-term approaches. Short-term treatment involves using eye drops like pilocarpine to lower intraocular pressure by increasing aqueous humor...
435
Open Angle Glaucoma: Treatment
391
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
391