Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Reciprocating Torsional Fatigue and Mechanical Tests of Thermal-Treated Nickel Titanium Instruments
Victor Talarico Leal Vieira1, Alejandro Jaime2, Carlos Garcia Puente2
1School of Dentistry, Grande Rio University (UNIGRANRIO), Rio de Janeiro, Brazil; Material Science Department. Military Institute of Engineering, Rio de Janeiro, Rio de Janeiro, Brazil; Department of Endodontics, Fluminense Federal University, Niteroi, Brazil.
Introduction:
This study aimed to evaluate the mechanical properties of Reciproc, Reciproc Blue, One RECI, and R-Motion, specifically focusing on their flexibility and buckling resistance and to assess their torsional fatigue behavior under 2 different reciprocation angles (150°/30° or 70°/30°).
Methods:
A total of 160 instruments (40 per brand) were tested. Flexion and buckling tests were conducted using a Universal testing machine (DL 200 MF). For the flexion test, instruments were positioned at a 45° angle and flexed toward the ground plane. In the buckling test, the load was applied along the instrument's axis. A 20 N load cell and a test speed of 15 mm/s were used for both tests. Torsion test was performed using a torsiometer. The test speed was 100 rpm and the instruments were driven by an endodontic motor in 150°/30° and 70°/30° reciprocating angles. One-way analysis of variance post hoc Tukey tests and Student's t test were used for statistical comparisons (α = 5%).
Results:
The instruments exhibited significant differences in flexibility (P < .05), with One RECI demonstrating the highest flexibility and Reciproc the lowest. In terms of buckling resistance, Reciproc showed the highest values, while One RECI had the lowest (P < .05). Reciproc and Reciproc Blue displayed greater torque to fracture compared to One RECI and R-Motion (P < .05). Reducing the reciprocation angle did not affect the torque to fracture (P > .05). However, the reduced angle of 70°/30° significantly increased the time to fracture for all instruments when compared to angle of 150°/30° (P < .05).
Conclusions:
The tested instruments demonstrated notable differences in their mechanical performance, highlighting variability in flexibility, buckling resistance, and torsional behavior. Reducing the reciprocation angle prolonged the time to fracture without affecting maximum torque, underscoring the influence of reciprocation kinematics on instrument performance.