Automated Measurement of Effective Radiation Dose by 18F-Fluorodeoxyglucose Positron Emission Tomography/Computed
Yujin Eom1, Yong-Jin Park2, Sumin Lee2
1Department of AI Mobility Engineering, Ajou University, Suwon 16499, Republic of Korea.
Background/Objectives:
Calculating the radiation dose from CT in 18F-PET/CT examinations poses a significant challenge. The objective of this study is to develop a deep learning-based automated program that standardizes the measurement of radiation doses.
Methods:
The torso CT was segmented into six distinct regions using TotalSegmentator. An automated program was employed to extract the necessary information and calculate the effective dose (ED) of PET/CT. The accuracy of our automated program was verified by comparing the EDs calculated by the program with those determined by a nuclear medicine physician (n = 30). Additionally, we compared the EDs obtained from an older PET/CT scanner with those from a newer PET/CT scanner (n = 42).
Results:
The CT ED calculated by the automated program was not significantly different from that calculated by the nuclear medicine physician (3.67 ± 0.61 mSv and 3.62 ± 0.60 mSv, respectively, p = 0.7623). Similarly, the total ED showed no significant difference between the two calculation methods (8.10 ± 1.40 mSv and 8.05 ± 1.39 mSv, respectively, p = 0.8957). A very strong correlation was observed in both the CT ED and total ED between the two measurements (r2 = 0.9981 and 0.9996, respectively). The automated program showed excellent repeatability and reproducibility. When comparing the older and newer PET/CT scanners, the PET ED was significantly lower in the newer scanner than in the older scanner (4.39 ± 0.91 mSv and 6.00 ± 1.17 mSv, respectively, p < 0.0001). Consequently, the total ED was significantly lower in the newer scanner than in the older scanner (8.22 ± 1.53 mSv and 9.65 ± 1.34 mSv, respectively, p < 0.0001).
Conclusions:
We successfully developed an automated program for calculating the ED of torso 18F-PET/CT. By integrating a deep learning model, the program effectively eliminated inter-operator variability.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Biological Effects of Radiation
Electronic Distance Measuring Instruments