Application of Online Multi-Internal Standard Calibration for Determination of Iodine by ICP-MS
Long-Hua Zhou1, Xing-Chen Lin1, Li-Li Suo1
1Physical and Chemical Department, The Collaboration Unit for Field Epidemiology of State Key Laboratory for Infectious Disease Prevention and Control, Nanchang Centre for Disease Control and Prevention, Nanchang, China.
Abstract:
This study presents a comprehensive evaluation of the application of online multi-internal standard calibration (M.ISC) in determining iodine concentrations through inductively coupled plasma mass spectrometry (ICP-MS). Notably, M.ISC streamlines the calibration process by requiring only a single standard solution, thereby enhancing sample throughput and minimizing liquid waste. In addition, unlike conventional internal standard (IS) methods, M.ISC omits the need for time-consuming species identification by utilizing multiple IS species simultaneously to minimize signal biases. The effectiveness of M.ISC was validated through the analysis of six standard reference samples, with the results of LOD and LOQ also being calculated by the error propagation approach. The traditional chemical analytical methods (TCAM), external standard calibration (EC) and single IS methods were also evaluated as comparative purpose. Nonetheless, M.ISC emerges as a straightforward matrix-correction strategy, offering a simple and efficient alternative to traditional calibration methods for iodine detection by ICP-MS.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Redox Titration: Iodimetry and Iodometry
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...