Validation and accuracy of the Hyfe cough monitoring system: a multicenter clinical study
Carlos Chaccour1,2,3, Isabel Sánchez-Olivieri4, Sarah Siegel5
1ISGlobal, Barcelona Institute for Global Health, Barcelona, Spain. carlos.chaccour@isglobal.org.
Background:
The ability to passively and continuously monitor coughing for prolonged periods of time would significantly improve cough management and research. To date there is no automated clinically validated cough monitor that can be routinely used in clinical care and research. Here we describe the validation of such an automated cough monitor.
Methods:
This multicenter observational study compared the results of the Hyfe CoughMonitor wrist-worn device with manually counted coughs in subjects with a variety of etiologies as they went about their usual daily activities. We collected 24 h of continuous sounds from subjects while they simultaneously wore a CoughMonitor and an audio recorder. Coughs were labelled by multiple trained annotators who listened to the continuous audio recordings using validated methodology. The time stamps of these human-detected coughs were compared to those of the CoughMonitor to determine the system's overall performance using event-to-event and hourly rate correlation analyses.
Results:
Over the 546 h monitored, 4,454 cough events were recorded; The overall sensitivity was 90.4% (95% CI of 88.3-92.2%). The overall false positive rate was 1.03 false positives per hour (95% CI of 0.84 to 1.24). The overall correlation between manual and CoughMonitor measured hourly coughing was high (Pearson correlation coefficient of 0.99). Two case studies of long-term monitoring of patients with chronic cough are presented.
Conclusion:
The present analysis of cough events demonstrated that the Hyfe CoughMonitor accurately reflects them with a high sensitivity and a low false positive rate. Future studies should focus on its potential role in the management of patients with cough in clinical practice.Registration Clinicaltrials.gov, NCT05723159.
Related Concept Videos
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...