Evaluation of mid vs distal left anterior descending artery measures in coronary physiology assessment

Fazal Dalal1, Yissela Escobedo2, Jose Emilio Exaire2

  • 1Department of Internal Medicine, Baylor Scott and White, Temple, TX, United States of America.

Insights

Coronary physiology assessment (CPA) wire positioning impacts results. Deeper placement in the left anterior descending artery (LAD) leads to lower coronary flow reserve (CFR) and higher index of microcirculatory resistance (IMR) in angina patients.

Area of Science:

  • Cardiology
  • Interventional Cardiology
  • Diagnostic Imaging

Background:

  • Angina with no obstructive coronary artery disease (ANOCA) affects ~40% of patients undergoing coronary angiography.
  • Coronary physiology assessment (CPA) is guideline-recommended for diagnosing ANOCA.
  • Optimal wire/sensor positioning for CPA via bolus coronary thermodilution is not established.

Purpose of the Study:

  • To investigate the impact of wire positioning on CPA outcomes.
  • To compare coronary flow reserve (CFR) and index of microcirculatory resistance (IMR) at different positions in the left anterior descending artery (LAD).

Main Methods:

  • Retrospective review of CPA data and demographics in ANOCA patients.
  • Evaluation of CFR and IMR with maximal hyperemia in the mid-LAD (4-6 cm) versus distal 2/3 LAD.
  • Statistical analysis using linear regression, paired t-test, and Bland-Altman analysis.

Main Results:

  • 42 patients (72% female, median age 60.5) were assessed.
  • Lower median CFR (2.71 vs 3.55) and higher median IMR (22.27 vs 16.41 mmHg*s) were observed in the distal LAD compared to the mid-LAD (p=0.01* for both).
  • Distal pressures showed no significant difference, but resting and hyperemic transit times differed significantly between mid and distal LAD segments.

Conclusions:

  • CPA outcomes (CFR, IMR) are significantly influenced by wire positioning.
  • Deeper wire placement in the LAD results in lower CFR and higher IMR values.
  • Standardized wire positioning is crucial for accurate CPA interpretation in ANOCA.
Abstract