Related Experiment Video
Updated: Jun 3, 2025

11:28
A Reference Broth Microdilution Method for Dalbavancin In Vitro Susceptibility Testing of Bacteria that Grow Aerobically
Published on: September 9, 2015
28.9K
Pooled Antibiotic Susceptibility Testing Performs Within CLSI Standards for Validation When Measured Against Broth
Emery Haley1, Frank R Cockerill2, Rick L Pesano2
1Department of Clinical Research, Pathnostics, Irvine, CA 92618, USA.
Antibiotics (Basel, Switzerland)
|January 8, 2025
Summary
A new pooled antibiotic susceptibility testing (P-AST) assay for urinary tract infections (UTIs) shows high accuracy and agreement with standard methods. This validated method meets Clinical and Laboratory Standards Institute (CLSI) guidelines for reliable antimicrobial susceptibility testing.
Area of Science:
- Microbiology
- Clinical Diagnostics
- Antimicrobial Resistance
Background:
- New antimicrobial susceptibility testing methods require validation against standard protocols for accuracy and error rates.
- Complicated and recurrent urinary tract infections (UTIs) necessitate reliable diagnostic tools.
Purpose of the Study:
- To validate a novel pooled antibiotic susceptibility testing (P-AST) assay for UTIs.
- To compare P-AST performance against broth microdilution (BMD) and disk diffusion (DD) using Clinical and Laboratory Standards Institute (CLSI) guidelines.
Main Methods:
- Consecutive clinical urine specimens were analyzed using standard urine culture, multiplex polymerase chain reaction (M-PCR), and P-AST.
- Antibiotic susceptibility testing (AST) was performed in parallel using P-AST, BMD, and DD.
- Analysis included assessment of heteroresistance for discrepant results.
Main Results:
- P-AST demonstrated essential agreement (EA%) ≥90% and met CLSI thresholds for very major errors (VMEs) <1.5% and major errors (MEs) <3.0% without heteroresistance correction.
- Categorical agreement (CA%) exceeded 88%, with most errors being minor.
- VME and ME rates decreased to <1.0% for P-AST when heteroresistance was accounted for.
Conclusions:
- The P-AST assay is validated within acceptable parameters according to CLSI criteria.
- P-AST shows comparable performance to BMD and DD for UTI diagnostics.
- The assay's accuracy is further improved by accounting for heteroresistance.
Keywords: