Lumped-Parameter and Finite Element Modeling of Heart Failure with Preserved Ejection Fraction
Published on: February 13, 2021
Exploring the influence of parametrized pulsatility on left ventricular washout under LVAD support: a computational
M R Schuster1, N Dirkes1, F Key2
1RWTH Aachen University, Aachen, Germany.
Optimizing pulsatility in left ventricular assist devices (LVADs) can improve washout and reduce complications. Short, intense pulses are most effective for left ventricular washout, with pulse timing being less critical.
Area of Science:
- Biomedical Engineering
- Cardiovascular Flow Dynamics
- Computational Fluid Dynamics
Background:
- Pulsatility in left ventricular assist devices (LVADs) is recognized for its potential to mitigate complications like blood stagnation and thrombosis.
- Determining the optimal pulsatile waveform for LVADs remains an active area of research, crucial for enhancing hemocompatibility and device efficacy.
Purpose of the Study:
- To develop and apply a computational framework for systematically quantifying the impact of parametrized pulsatility on washout within a left ventricle.
- To investigate the effects of different pulsatile flow rate parameters on the efficiency of ventricular washout.
Main Methods:
- A framework integrating parametrized full-order simulations, reduced-order models, and sensitivity analysis was employed.
- An idealized 2D left ventricle model was used with a parametrized sinusoidal LVAD flow rate to simulate washout dynamics.
- The framework enabled significant computational speed-ups in the sensitivity analysis process.
Main Results:
- The study identified that short, intense pulsatile flows provide the most effective washout in the idealized left ventricle.
- The duration between consecutive pulses was found to have a negligible impact on the overall washout efficiency.
- The developed framework demonstrated efficiency gains proportional to the number of samples needed for sensitivity analysis.
Conclusions:
- The findings suggest that optimizing pulse characteristics, specifically intensity and duration, is key for improving LVAD performance in terms of washout.
- This research provides a systematic approach to analyzing pulsatility effects, offering valuable insights for the design of next-generation LVADs.
- The computational framework presented can accelerate the optimization process for LVAD flow dynamics.