Evaluation of point-of-care haemoglobin measurement accuracy in surgery (PREMISE) and implications for transfusion
Karine Brousseau1, Leah Monette1, Daniel I McIsaac2
1Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada; Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
Background:
Point-of-care testing devices to measure haemoglobin (Hgb) frequently inform transfusion decision-making in surgery. This study aimed to examine their accuracy in surgery, focusing on Hgb concentrations of 60-100 g L-1, a range with higher potential for transfusion.
Methods:
This was a prospective diagnostic cohort study focused on method comparison, conducted at two academic hospitals. Consecutive patients undergoing noncardiac surgery and requiring point-of-care Hgb measurements were eligible. Hgb concentrations from arterial and central venous blood samples were measured concurrently using three devices and compared with laboratory Hgb. The primary outcome was individual pairwise comparisons between point-of-care and laboratory Hgb values; agreement was determined based on a threshold of within 4 g L-1. The primary analysis consisted of computing limits of agreement.
Results:
A total of 1735 intraoperative blood samples were collected (1139 participants); 680 samples had a laboratory Hgb <100 g L-1. The limits of agreement among those with Hgb <100 g L-1 were -9.5 to 8.0 g L-1 for HemoCue®, -16.2 to 11.5 g L-1 for i-STAT®, and -14.7 to 40.5 g L-1 for Rad-67®. HemoCue was associated with a 5.8% incidence of potentially clinically significant transfusion error, whereas i-STAT and Rad-67 were associated with 25.3% and 28.2%, respectively. HemoCue yielded Hgb measurements within 10 g L-1 in 98% of intraoperative blood samples.
Conclusions:
No point-of-care Hgb device demonstrated limits of agreement that were smaller than the agreement difference of 4 g L-1. Despite this, HemoCue can be safely used to inform transfusion decisions in surgery, given its error probability of <4% in transfusion scenarios.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...