Validation of Aerobic Capacity (VO2max) and Pulse Oximetry in Wearable Technology
Bryson Carrier1, Sofia Marten Chaves1, James W Navalta1
1School of Integrated Health Sciences, Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, NV 89154, USA.
Introduction:
As wearable technology becomes increasingly popular and sophisticated, independent validation is needed to determine its accuracy and potential applications. Therefore, the purpose of this study was to evaluate the accuracy (validity) of VO2max estimates and blood oxygen saturation measured via pulse oximetry using the Garmin fēnix 6 with a general population participant pool.
Methods:
We recruited apparently healthy individuals (both active and sedentary) for VO2max (n = 19) and pulse oximetry testing (n = 22). VO2max was assessed through a graded exercise test and an outdoor run, comparing results from the Garmin fēnix 6 to a criterion measurement obtained from a metabolic system. Pulse oximetry involved comparing fēnix 6 readings under normoxic and hypoxic conditions against a medical-grade pulse oximeter. Data analysis included descriptive statistics, error analysis, correlation analysis, equivalence testing, and bias assessment, with the validation criteria set at a concordance correlation coefficient (CCC) > 0.7 and a mean absolute percentage error (MAPE) < 10%.
Results:
The Garmin fēnix 6 provided accurate VO2max estimates, closely aligning with the 15 s and 30 s averaged laboratory data (MAPE for 30 s avg = 7.05%; Lin's concordance correlation coefficient for 30 s avg = 0.73). However, it failed to accurately measure blood oxygen saturation (BOS) under any condition or combined analysis (MAPE for combined conditions BOS = 4.29%; Lin's concordance correlation coefficient for combined conditions BOS = 0.10).
Conclusion:
While the Garmin fēnix 6 shows promise for estimating the VO2max, reflecting its utility for both individuals and researchers, it falls short in accurately measuring BOS, limiting its application for monitoring acclimatization and managing pulmonary diseases. This research underscores the importance of validating wearable technology to leverage its full potential in enhancing personal health and advancing public health research.
More Related Videos
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...