Low-frequency response characteristics of three Grass model 7 polygraphs.

D D Hickey, J Zaharkin

    Journal of Applied Physiology (Bethesda, Md. : 1985)
    |March 1, 1985
    PubMed
    Summary

    Low-frequency analysis of Grass model 7 polygraphs revealed errors exceeding manufacturer specifications above 6 Hz. Pen-oscillograph mechanics, including inertia and friction, were identified as primary error sources, with proposed solutions to mitigate inaccuracies.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    A comparison of respiratory function in divers breathing with a mouthpiece or a full face mask.

    Undersea biomedical research·1987
    Same author

    Simulated breath-hold diving to 20 meters: cardiac performance in humans.

    Journal of applied physiology (Bethesda, Md. : 1985)·1987
    Same author

    Respiratory function in the upright, working diver at 6.8 ATA (190 fsw).

    Undersea biomedical research·1987
    Same author

    A simple, inexpensive device for aspiration biopsy.

    Surgery, gynecology & obstetrics·1987
    Same author

    A simple device for the direct measurement of mean arterial pressure and for calibration of arterial pressure monitors.

    Journal of medical engineering & technology·1986
    Same author

    Cardiac performance in humans during breath holding.

    Journal of applied physiology (Bethesda, Md. : 1985)·1986

    Area of Science:

    • Biomedical Engineering
    • Physiological Monitoring

    Background:

    • Accurate physiological signal recording is crucial for diagnosis.
    • Polygraphs are essential tools in clinical and research settings.
    • Understanding instrument limitations, especially at low frequencies, is vital.

    Purpose of the Study:

    • To evaluate the low-frequency response of Grass model 7 polygraphs.
    • To identify sources of error in polygraph recordings.
    • To propose methods for reducing measurement inaccuracies.

    Main Methods:

    • Performed low-frequency response analysis on three Grass model 7 polygraphs.
    • Measured signal error across various frequencies (0.1-6+ Hz).
    • Assessed the influence of frequency, pen deflection, and paper speed on error magnitude.

    Related Experiment Videos

    Main Results:

    • Errors were within the manufacturer's range (+5 to -10%) for human respiration frequencies (0.1-3 Hz).
    • Significant errors occurred at frequencies above 6 Hz under specific conditions.
    • Error magnitude correlated positively with frequency and negatively with pen deflection amplitude and paper speed.
    • Pen-oscillograph apparatus (inertia, friction) identified as the main source of low-frequency error.

    Conclusions:

    • Grass model 7 polygraphs exhibit acceptable low-frequency performance for typical respiration monitoring.
    • Higher frequencies and specific operational parameters can lead to substantial recording errors.
    • The pen-oscillograph system is a key contributor to low-frequency inaccuracies.
    • Strategies to minimize pen-related friction and inertia can improve polygraph accuracy.