A comparative analysis of current С-peptide assays compared to a reference method: can we overcome inertia to
Curt Rohlfing1, Gregory Petroski1, Shawn M Connolly1
1Pathology & Anatomical Sciences, University of Missouri, Columbia MO, USA.
Objectives:
C-peptide is an equimolar by-product of insulin biosynthesis. It is used clinically to assess insulin secretion and differentiate types of diabetes. However, the lack of standardization across assays limits its broader application. This study aimed to examine discrepancies between the leading C-peptide measurement methods used in clinical laboratories and propose a solution to reduce them based on a complete traceability chain.
Methods:
Two sets of serum samples were distributed to 10 manufacturers of C-peptide assays. The first set (A, n=20) was analyzed independently by each manufacturer, who then returned their results to us. Subsequently, we sent out the second set (B, n=20) along with the reference values for set A. For set B, each manufacturer provided both non-calibrated and recalibrated values for each sample. The recalibration was performed according to each manufacturer's internal standard protocols. We assessed how recalibration affected agreement between methods and alignment with the reference method. Non-parametric statistical approaches, including Passing-Bablok regression, level of agreement, and standard deviation analysis, were applied to compare data from multiple perspectives.
Results:
Despite most manufacturers using the same WHO C-peptide calibrator material, significant disagreement was observed between methods prior to recalibration. Recalibration with matrix-appropriate serum samples reduced the discordance among assays, bringing them closer to the reference method. Overall, recalibration reduced both systematic bias and individual assay disagreement.
Conclusions:
These findings underscore the importance of appropriate calibration schemes to improve agreement across C-peptide assays, enhancing the accuracy of C-peptide testing for clinical practice.
Related Concept Videos
Decision Making: Traditional Method
First, a specific claim about the population parameter is decided based on the research question and is stated in a simple form. Further, an opposing statement to this claim is also stated. These statements can act as null and alternative hypotheses, out of which a null hypothesis would be a...
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Introduction to Nonparametric Statistics
One of...
Comparing Experimental Results: Student's t-Test
Statistical Methods to Analyze Parametric Data: Student t-Test and Goodness-of-Fit Test
The Student's t-test is a statistical test that examines if there is a statistically significant difference between the means of two groups. This test is instrumental when dealing with data...