Systematic evaluation of adhesives for implant fixation in multimodal functional brain MRI
Anna Zsófia Szinyei1, Bastian Maus1, Jonas Q Schmid2
1Translational Research Imaging Center (TRIC), Clinic of Radiology, University of Münster, Albert-Schweitzer-Campus 1, building A16, 48149, Münster, Germany.
Objective:
Invasive multimodal fMRI in rodents is often compromised by susceptibility artifacts from adhesives used to secure cranial implants. We hypothesized that adhesive type, shape, and field strength significantly affect susceptibility artifacts, and systematically evaluated various adhesives.
Materials And Methods:
Thirty-one adhesives were applied in constrained/unconstrained geometries and imaged with T2*-weighted EPI at 7.0 and 9.4 T to assess artifact depths. Spherical and flat patch shapes, both unconstrained geometries, were compared for artifact depth in vitro and in vivo. Adhesion strength was assessed on post-mortem mouse crania. Finally, an integrative scoring system rated adhesive properties, including artifact depth, handling, and adhesion strength.
Results:
Susceptibility artifacts were two times larger at 9.4 than at 7.0 T (p < 0.001), strongest at the patch edges, and deeper with spherical than flat patches (p < 0.05). Artifact size depended more on shape and volume after curing than adhesive type. Our integrative scoring system showed resins, bonding agents, and acrylics offered the best overall properties, while silicones and cements were less favorable.
Discussion:
Adhesive selection requires balancing handling, curing time, strength, and artifact depth. To minimize artifacts, adhesives should be applied in a spread-out, flat and thin layer. Our integrative scoring system supports classification of future classes of adhesives.
More Related Videos
08:23A Multimodal Imaging- and Stimulation-based Method of Evaluating Connectivity-related Brain Excitability in Patients with Epilepsy
Published on: November 13, 2016
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies for Cardiovascular System IV: CMRI