The proton resonance enhancement for CEST imaging and shift exchange (PRECISE) family of RF pulse shapes for CEST MRI
Zinia Mohanta1,2, Julia Stabinska1,2, Assaf A Gilad3,4
1Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
Purpose:
To optimize a 100 ms pulse for producing CEST MRI contrast and evaluate in mice.
Methods:
A gradient ascent algorithm was employed to generate a family of 100 point, 100 ms pulses for use in CEST pulse trains (proton resonance enhancement for CEST imaging and shift exchange). Gradient ascent optimizations were performed for exchange rates = 500, 1500, 2500, 3500, and 4500 s-1; and labile proton offsets (Δω) = 9.6, 7.8, 4.2, and 2.0 ppm. Seven proton resonance enhancement for CEST imaging and shift exchange pulse shapes were tested on an 11.7 T scanner using a phantom containing three representative CEST agents with peak saturation B1,peak = 4 μT. The pulse producing the most contrast in phantoms was then evaluated for CEST MRI pH mapping of the kidneys in healthy mice after iopamidol administration.
Results:
The most promising pulse in terms of contrast performance across all three phantoms was the 9.6 ppm, 2500 s-1 optimized pulse with ˜2.7 × increase in asymmetric magnetization transfer ratio (MTRasym) over Gaussian, and ˜ 1.3 times over Fermi pulses for the same B1,peak = 4 μT. This pulse also displayed a large improvement in contrast over the Gaussian pulse after administration of iopamidol in live mice.
Conclusion:
A new 100-ms pulse was developed based on gradient ascent optimizations, which produced better contrast compared to standard Gaussian and Fermi pulses in phantoms. This shape also showed a substantial improvement for CEST MRI pH mapping in live mice over the Gaussian shape and appears promising for a wide range of CEST applications.
More Related Videos
08:01Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
Published on: September 26, 2016
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction