Comparative image quality and dosimetric performance of two generations of dedicated breast CT systems
Juan J Pautasso1, Camille D E Van Speybroeck1, Koen Michielsen1
1Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
Background:
Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.
Purpose:
To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.
Methods:
The imaging performance was evaluated through key metrics: modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) in the projection domain. In the image domain, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and the visibility of calcifications were analyzed using a quality control (QC) phantom with masses and calcification clusters. Air kerma and tube output were measured and mean glandular dose (MGD) estimated for different phantom sizes for dosimetric characterization of the acquisition protocols set by the automatic exposure control (AEC).
Results:
GEN2 outperformed GEN1 at higher spatial frequencies, with 57% of the MTF observed at 1 cycles/mm compared to 43% for GEN1. For a 2 mm diameter mass, GEN2 showed 60% higher CNR and 63% higher SNR. However, for larger masses, GEN1 outperformed GEN2, with CNR and SNR values higher by 12% to 44% and 14% to 43%, respectively. GEN2 also achieves higher DQE across the frequency spectrum, with 45% at 1 cycle/mm, compared to GEN1's 20%. Regarding calcifications in the QC phantom, the 320 µm calcifications resulted in distinct full-width-at-half-maxima (FWHM ± SD), with 897 ± 58 µm for GEN1 and 811 ± 127 µm for GEN2, with a p-value of 0.19. For 290 µm calcifications, GEN1's FWHM was 866 ± 129 µm, while GEN2's was narrower at 665 ± 57 µm, with a p-value of 0.01. The tube output was higher for GEN1 (45.2 mGy/mAs) compared to GEN2 (31.5 mGy/mAs). Additionally, GEN2 resulted in 8% lower MGD values compared to GEN1.
Conclusion:
While GEN1 offers better CNR and SNR for larger masses, GEN2 provides superior resolution for calcifications, better MTF, improved DQE, and lower MGD at AEC-determined settings.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...