Effect of different scanning threshold triggers on the image quality of brain computed tomography angiography: a
Jun Xu1,2,3, Jian Shen4, Qian Dong5
1Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Background:
The scanning trigger threshold affects image quality. The aim of this study was to investigate the effect of different scanning trigger thresholds on brain computed tomography angiography (CTA) image quality.
Methods:
In this prospective study, 80 patients undergoing brain CTA examinations with dual-layer CT (DLCT) were randomly divided into group A and group B, with 40 patients in each group. In group A, the CT value of the internal carotid artery at the level of the fourth cervical vertebra was monitored, and the scan was initiated once the CT value reached 100 Hounsfield units (HU). In group B, the trigger threshold was set at 60 HU, with all other parameters kept consistent with those of group A. Finally, the image quality of the 50-keV virtual monoenergetic images (VMIs) was evaluated, including the CT values of the internal carotid artery (CTICA), middle cerebral artery (CTMCA), sinus confluence (CTSC), cerebral white matter (CTCWM), background noise (BN), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and subjective scoring.
Results:
All images met the diagnostic imaging requirements. Group B showed significantly lower values than did group A for CTICA (371.97±51.81 vs. 442.64±83.39 HU), CTMCA (345.80±50.72 vs. 405.87±82.81 HU), CTSC (90.44±21.30 vs. 138.87±37.37 HU), CTCWM (31.98±5.66 vs. 38.86±5.68 HU), SNR (108.64±21.05 vs. 126.79±30.87), and CNR (98.58±19.72 vs. 114.65±29.56) (all P values <0.05) but the value for BN was not significantly different (P>0.05). However, the subjective scores in group B were significantly higher than those in group A (χ2=19.013; P<0.05).
Conclusions:
For brain CTA imaging in DLCT (50 keV VMIs), lowering the scan trigger threshold to 60 HU helped to reduce venous artifacts and improve image quality (as evidenced by improved subjective scores) and also suggests the potential for a further reduction of the contrast dose.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...