Related Experiment Video
Updated: May 5, 2026

Colorimetric Paper-based Detection of Escherichia coli, Salmonella spp., and Listeria monocytogenes from Large Volumes of Agricultural Water
Published on: June 9, 2014
Enhancing Biocide Safety of Milk Using Biosensors Based on Cholinesterase Inhibition
Lynn Mouawad1,2, Georges Istamboulie1,2, Gaëlle Catanante1,2
1Biosensors Analysis Environment Group (BAE-LBBM), Université de Perpignan, Via Domitia, 52 Avenue Paul Alduy, Cedex, F-66860 Perpignan, France.
This study developed an electrochemical biosensor for detecting benzalkonium chloride (BAC) and didecyldimethylammonium chloride (DDAC). A liquid-liquid extraction method was crucial for accurate detection in milk samples, meeting European regulations.
Area of Science:
- Electrochemistry
- Biosensor Technology
- Analytical Chemistry
Background:
- Benzalkonium chloride (BAC) and didecyldimethylammonium chloride (DDAC) are widely used biocides in the agri-food sector.
- Accurate detection of these biocides is essential for ensuring food safety and regulatory compliance.
- Existing detection methods may face challenges with complex sample matrices like milk.
Purpose of the Study:
- To develop and validate a sensitive electrochemical biosensor for BAC and DDAC detection.
- To assess the biosensor's performance in a relevant matrix (milk).
- To establish a reliable method for monitoring biocide levels in dairy products.
Main Methods:
- Immobilization of acetylcholinesterase (DM AChE) and butyrylcholinesterase (BChE) on screen-printed electrodes.
- Development of an electrochemical detection strategy for BAC and DDAC.
- Implementation of a liquid-liquid extraction technique for sample preparation.
- Validation of the biosensor against European regulatory limits.
Main Results:
- Achieved low limits of detection (LODs) in buffer: 0.26 µM for BAC (BChE) and 0.04 µM for DDAC (DM AChE).
- Observed a significant decrease in sensitivity when testing directly in milk samples.
- Demonstrated that liquid-liquid extraction restored sensor sensitivity and accuracy in milk.
- Confirmed compliance with European Union regulations for dairy products (max 0.28 µM).
Conclusions:
- The developed electrochemical biosensor shows promise for biocide detection.
- Sample pretreatment (liquid-liquid extraction) is critical for accurate analysis in milk.
- The method provides a reliable tool for monitoring BAC and DDAC in the agri-food industry, meeting regulatory standards.
More Related Videos
08:06The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions