Characterization of LIPUS Parameters Suitable for Hip Bone Fracture
Poornima Palanisamy1, Shuai Li1, Queenie Tsung-Kwan Shea1
1Department of Biomedical Engineering, The Hong Kong Polytechnic University, Hong Kong SAR, China.
Ultrasound in Medicine & Biology
|January 28, 2025
Summary
Low-intensity pulsed ultrasound (LIPUS) at 0.5 MHz and a consistent intensity of 30 mW/cm² significantly enhances hip bone fracture healing in rabbits. This frequency demonstrates superior bone healing compared to 1.5 MHz LIPUS.
Area of Science:
- Biomedical Engineering
- Orthopedic Surgery
- Regenerative Medicine
Background:
- Hip bone fractures represent a significant clinical challenge, impacting mobility and quality of life.
- Ultrasound therapy, particularly low-intensity pulsed ultrasound (LIPUS), has emerged as a promising non-invasive treatment modality for bone healing.
- Optimizing LIPUS parameters, such as frequency and intensity, is crucial for maximizing therapeutic efficacy.
Purpose of the Study:
- To evaluate the efficacy of different low-intensity pulsed ultrasound (LIPUS) frequencies (0.5 MHz and 1.5 MHz) for promoting hip bone fracture healing.
- To determine the optimal LIPUS frequency and intensity for enhanced bone regeneration at the fracture site.
- To investigate the impact of acoustic attenuation in soft tissues on ultrasound intensity delivery.
Main Methods:
- Ex vivo acoustic attenuation measurements were performed using human cadaver and porcine tissues to assess ultrasound propagation.
- In vivo study involved 20 hip-fractured rabbits divided into control and LIPUS treatment groups (0.5 MHz and 1.5 MHz).
- A compensation method was employed to maintain a constant intensity of 30 mW/cm² at the fracture site, accounting for tissue attenuation.
Main Results:
- Ultrasound intensity significantly decreased with tissue depth; 1.5 MHz intensity dropped to 6.16 mW/cm² through 5.0 cm of tissue.
- The compensation method successfully maintained 30 mW/cm² intensity at the fracture site for both 0.5 MHz and 1.5 MHz groups.
- The 0.5 MHz LIPUS group demonstrated superior outcomes in bone histomorphometry, mechanical properties, and histological evaluation compared to the 1.5 MHz and control groups.
Conclusions:
- A LIPUS frequency of 0.5 MHz, delivered at a consistent intensity of 30 mW/cm² to the fracture site, is more effective for treating hip bone fractures.
- The findings suggest that 0.5 MHz LIPUS offers a promising therapeutic strategy for accelerating bone healing in hip fractures.
- Maintaining consistent ultrasound intensity at the fracture site through compensation methods is critical for effective LIPUS treatment.
Keywords:
Animal studyHip bone fracturesLow-intensity pulsed ultrasound stimulation (LIPUS)Optimal parameters