Assessment of a New CT Detector and Filtration Technology: Part 2-Image Quality in Phantoms, Cadavers, and Patients
Izabella Barreto1, Graham Stoddard, Anahita Heshmat
1Department of Radiology, College of Medicine, University of Florida, Gainesville, FL.
Purpose:
The purpose of this work was to evaluate the image quality of a commercial CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO).
Methods:
CT number, noise, contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectrum (NPS) were assessed using the ACR CT Accreditation phantom scanned with various acquisitions at 80 kV, 100 kV, 120 kV, and 135 kV, each with multiple CTDI vol values of 20 mGy, 40 mGy, and 65 mGy. Artifacts were evaluated in an anthropomorphic head phantom, a cadaver head, and in patient studies. Two neuroradiologists assessed image quality features in various patients who were examined with unenhanced brain CT on both scanners.
Results:
Compared with the conventional scanner, for the same CTDI vol , the PVO scanner produced 20.3% less image noise ( P < 0.001), 18.9% higher CNR ( P < 0.01), and 24.6% higher spatial resolution ( P < 0.001). Streak artifacts were less severe with the PVO scanner for the phantom, cadaver, and patient scans ( P < 0.05). Radiologists scored the PVO scanner as significantly better for visualization of the cerebrospinal fluid space over the cerebral sulci in high convexity, image noise in gray and white matter, and artifacts in the posterior fossa. They also significantly preferred the PVO scanner for visualization of the border between brain gray and white matter, cerebrospinal fluid space around the mesencephalon, and overall diagnostic acceptability.
Conclusions:
For matched CTDI vol values, the scanner equipped with PVO technology produced better objective and subjective image quality metrics in brain CT imaging compared with a conventional CT scanner without PVO. In clinical settings, PVO may allow for lower doses while enhancing imaging through dense areas, improving visualization of subtle details, and offering more effective options for examining obese patients.This research received financial support from Canon Medical Systems USA. The study design and data were fully controlled by the coauthors, of which none are employees or consultants of Canon Medical Systems USA.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET