Depth-electrode stimulation and concurrent functional MRI in humans: Factors influencing heating with body coil
Hiroyuki Oya1, Ralph Adolphs2, Matthew A Howard3
1Department of Neurosurgery, University of Iowa, Iowa City, IA, USA; Iowa Neuroscience Institute, University of Iowa, Iowa City, IA, USA.
Abstract:
Electrical-stimulation fMRI (es-fMRI) combines direct stimulation of the brain via implanted electrodes with simultaneous rapid functional magnetic resonance imaging of the evoked response. Widely used to map effective functional connectivity in animal studies, its application to the human brain has been limited due to safety concerns. In particular, the method requires reliable prediction and minimization of local tissue heating close to the electrodes, which will vary with imaging parameters and hardware configurations. Electrode leads for such experiments typically remain connected to stimulators outside the magnet room and cannot therefore be treated as electrically short at the radio frequencies employed for 1.5 T and 3 T fMRI. The potential for significant absorption and scattering of radiofrequency energy from excitation pulses during imaging is therefore a major concern. We report a series of temperature measurements conducted in human brain phantoms at two independent imaging centers to characterize factors effecting RF heating of electrically long leads with body coil transmission at 3 Tesla for temporal RMS RF transmit fields ( [Formula: see text] ) up to 3.5 µT including multiband echo planar imaging and 3D T2w turbo spin echo imaging. Under all conditions tested, with one exception, the temperature rise measured immediately adjacent to electrode contacts in a head-torso phantom with body coil RF transmission was less than 0.75 °C. We provide detailed quantification across a range of configurations and conclude with specific recommendations for cable routing that will help ensure the safety of es-fMRI in humans and provide essential data to institutional review boards.
More Related Videos
07:42Combined Transcranial Magnetic Stimulation and Electroencephalography of the Dorsolateral Prefrontal Cortex
Published on: August 17, 2018
05:26Author Spotlight: Low-Cost Electroencephalographic Recording System Combined with a Millimeter-Sized Coil to Transcranially Stimulate the Mouse Brain In Vivo
Published on: May 26, 2023
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging