Immediate Effects of Goldmann Applanation Tonometry on Central Corneal Thickness Measurements Using Contact and
Adiraj S Sibia1, Raphael G Banoub2, Isabel Eaddy3
1Department of Ophthalmology, Florida Atlantic University-Broward Health North, Fort Lauderdale, USA.
Cureus
|February 3, 2025
Summary
Goldmann Applanation Tonometry (GAT) may affect central corneal thickness (CCT) measurement reliability, particularly with contact ultrasound pachymetry. This impacts diagnostic accuracy for ocular conditions.
Area of Science:
- Ophthalmology
- Corneal Physiology
- Diagnostic Imaging
Background:
- Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- Goldmann Applanation Tonometry (GAT) is the gold standard for IOP measurement.
- Central corneal thickness (CCT) influences IOP readings and is vital for accurate diagnosis.
Purpose of the Study:
- To assess the immediate effect of GAT on CCT measurement reliability.
- To compare CCT measurements between contact and non-contact pachymetry methods post-GAT.
Main Methods:
- Fifty-one participants underwent serial CCT measurements using ultrasound pachymetry (USP), optical low-coherence reflectometry (OLCR), and anterior segment optical coherence tomography (AS-OCT).
- Measurements were taken at baseline and after GAT over three days.
- Statistical analysis included ANOVA and paired t-tests to compare mean CCT and standard deviation.
Main Results:
- Mean CCT showed no significant change after GAT across all modalities.
- Contact USP yielded consistently thicker CCT readings than non-contact methods (OLCR, AS-OCT).
- Contact USP demonstrated a significant increase in CCT standard deviation and reduced repeatability post-GAT.
Conclusions:
- Contact IOP measurement techniques like GAT may compromise the reliability and repeatability of CCT measurements.
- Increased CCT standard deviation post-GAT warrants consideration for accurate diagnosis and treatment planning.
- Further research is recommended due to the study's limited sample size.
Keywords:
central corneal thicknessgoldmann applanation tonometryophthalmology clinical practicepachymetryquality measurementMore Related Videos
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
648
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
648
Open Angle Glaucoma: Treatment
387
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
387
Angle Closure Glaucoma: Treatment
432
Angle-closure glaucoma, or closed-angle glaucoma, is an eye condition where the iris bulges out and blocks the iridocorneal angle, resulting in a buildup of aqueous humor and increased intraocular pressure. Immediate medical attention is necessary due to the sudden onset of symptoms. The treatment for angle-closure glaucoma includes short-term and long-term approaches. Short-term treatment involves using eye drops like pilocarpine to lower intraocular pressure by increasing aqueous humor...
432