Large Single Center Comparison of Novel and Reusable Duodenoscopes Shows Similar Contamination Rates That Do Not
Sriya S Muralidharan1, Divya Chalikonda2, Grace E Kim2
1Section of Gastroenterology, Hepatology, and Nutrition, Department of Internal Medicine, University of Chicago, Chicago, Illinois, USA.
Novel duodenoscopes with disposable caps and reusable duodenoscopes show similar bacterial contamination rates. This finding supports consistent high-level disinfection practices for all duodenoscope types.
Area of Science:
- Gastroenterology
- Infectious Disease Control
- Medical Device Safety
Background:
- Limited data exists comparing contamination rates between reusable and novel duodenoscopes.
- The clinical impact of bacterial contamination in novel duodenoscopes remains unclear.
- This study addresses the need for comparative data on duodenoscope contamination.
Purpose of the Study:
- To evaluate and compare the pathogenic bacterial contamination rates of novel duodenoscopes with disposable caps and reusable duodenoscopes.
- To assess the clinical implications for patients exposed to contaminated duodenoscopes.
Main Methods:
- An observational study was conducted to assess bacterial contamination in reusable and novel duodenoscopes.
- Positive cultures from duodenoscopes led to patient chart reviews for blood culture data and sepsis indicators.
- Sepsis was evaluated using quick sequential organ failure assessment and systemic inflammatory response syndrome criteria.
Main Results:
- The contamination rate for reusable duodenoscopes was 0.80%, while novel duodenoscopes showed 0.0% contamination.
- No statistically significant difference in contamination rates was observed between the two types of duodenoscopes (P = 0.141).
- Gram-negative rods were identified in 4 positive cultures from reusable duodenoscopes; however, no exposed patients developed infection or sepsis.
Conclusions:
- Bacterial contamination rates are similar for novel duodenoscopes with disposable caps and reusable duodenoscopes.
- The low contamination rates observed support the efficacy of current high-level disinfection practices.
- Consistent adherence to disinfection protocols is crucial regardless of duodenoscope generation.