Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
[Accuracy of Proton Density Fat Fraction Measurement Using Chemical Shift-encoded MRI with Fast Imaging Techniques]
Tomofumi Misaka1,2, Satoshi Takenaka1, Takayuki Ishida2
1Department of Radiology, Kindai University Nara Hospital.
Purpose:
To investigate the accuracy of proton density fat fraction (PDFF) measurement using chemical shift-encoded MRI (CSE-MRI) with fast imaging techniques in a phantom.
Methods:
A 1.5T imaging system (Prodiva; Philips Healthcare) and PDFF phantom (Fat Fraction Phantom Model 300; Calimetrix) were used in this study. The acquisitions without fast imaging techniques (conventional acquisition), with parallel imaging in phase-encode direction (SENSE acquisition), with compressed sensing (CS-SENSE acquisition), and with parallel imaging in both phase-encode and slice-encode direction (Dual-SENSE acquisition) were performed. The following acceleration factors in SENSE and CS-SENSE acquisition were used: 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, and 8.0. For Dual-SENSE acquisition, the same acceleration factors (1.5, 2.0, 3.0, 4.0, 5.0) were set in each of the two directions. The relationships between reference PDFF values and PDFF measurements obtained using each acquisition were assessed using linear regression analysis and Bland-Altman analysis.
Results:
According to the linear regression analysis, the slopes and intercepts of regression lines were from 0.87 to 1.02 and from 0.06% to 3.55%, respectively. According to Bland-Altman analysis, there were fixed bias between reference PDFF values and PDFF measurements obtained using SENSE acquisition with reduction factor 8.0 and Dual-SENSE acquisition with reduction factor 5.0. For CS-SENSE acquisition with reduction factor from 7.0 to 8.0, SENSE acquisition with reduction factor from 3.0 to 8.0, and Dual-SENSE acquisition with reduction factor from 2.0 to 5.0, some vials had ±1.5% or more errors between the reference PDFF values and PDFF measurements in the range of 0% to 50% PDFF.
Conclusion:
In CS-SENSE acquisition, the accuracy of PDFF measurement was maintained within 1.5% up to a reduction factor 6.0. The accuracy of PDFF measurement was maintained within 1.5% up to a reduction factor 2.0 in SENSE acquisition and a reduction factor 1.5 in Dual-SENSE acquisition.
Related Concept Videos
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Magnetic Resonance Imaging