Fast and accurate peak skin dose estimation method for interventional fluoroscopy patients
Haegin Han1, Seth W Streitmatter2, Cari M Kitahara1
1Radiation Epidemiology Branch, Division of Cancer Epidemiology and Genetics, National Cancer Institute, National Institute of Health, Rockville, Maryland, USA.
Medical Physics
|February 6, 2025
Summary
A new method combines Monte Carlo simulation with noise reduction to accurately calculate peak skin dose (PSD) and skin dose distributions during fluoroscopically guided interventions (FGIs). This approach is significantly faster than traditional methods, offering a practical solution for radiation safety in medical imaging.
Area of Science:
- Medical Physics
- Radiological Sciences
- Computational Imaging
Background:
- Accurate peak skin dose (PSD) estimation is critical in fluoroscopy and fluoroscopically guided interventions (FGIs) to prevent radiation-induced skin injuries.
- Current analytical methods for PSD calculation have limitations due to simplified assumptions about X-ray beam conditions, patient geometry, and positioning.
- Full Monte Carlo (MC) simulations offer high accuracy but are computationally intensive and time-consuming, limiting their clinical applicability.
Purpose of the Study:
- To develop an efficient and accurate method for calculating PSD and skin dose distributions by integrating MC simulation with noise reduction techniques.
- To provide a practical alternative to existing methods for both clinical and academic use in radiation dosimetry.
Main Methods:
- A two-step approach combining a rough MC simulation with iterative noise reduction was developed for PSD and skin dose distribution calculation.
- The methodology's performance was validated across six fluoroscopy scenarios.
- Results were benchmarked against full MC simulations, considered the gold standard in radiation dosimetry.
Main Results:
- The novel method accurately estimates PSD and skin dose distributions, achieving discrepancies within 3% for PSD and less than 10% for voxel-wise skin dose compared to full MC simulations.
- The iterative noise reduction effectively handles statistical errors up to 20% in rough MC simulations.
- Computation time was reduced to seconds on a personal computer, representing at least a 10^4-fold speed improvement over full MC simulations.
Conclusions:
- The developed method offers rapid and accurate calculation of PSD and skin dose distribution, proving valuable for research and clinical applications.
- Integration into the National Cancer Institute Dosimetry System for Radiography and Fluoroscopy (NCIRF) will improve accessibility.
- Future enhancements, including a phantom library, will enable patient-specific dose assessments for personalized radiation safety.